HYG-20ADC Resistance Tester

RELEVANT RECOMMENDATIONS
-
VIDEO CONTENTS
1. Wiring: Connect the tested product with the test wiring post of the machine through special cables, which is firmly connected, and connect the ground wire well at the same time.
2. Current selection: Open the power switch (I is on, O is off) and the maximum current value 20A will be displayed on the display screen. At this time, the preset current of the tested product can be selected by selecting the key. Every time the selection key is pressed, the current values will roll on the display screen, 5A, 10A, 20A.
3. Test: After choosing the current, press the confirmation key and start the test. The head of the meter indicates the selected current value at the same time. When the confirmation key is pressed, the display screen shows "charging" and after a few seconds, shows "testing", which means that the charging has been completed. After entering the test state, the measured resistance value will be displayed in a few seconds. (The screen will display both the selected current value and the measured resistance value, I=""A"; R=""m_/_)
4. After testing, press the "reset" button, the power supply of the instrument will be disconnected from the winding, discharge at the same time, sound alarm, ammeter will return to zero, then the display screen will return to the initial state, after discharging the sound, it can be reconnected for the next measurement or disconnection of the test line and the power line end measurement.
-
TEST EQUIPMENT LISTS
Direct Resistance Instrument Main Engine
One station
Model 20A Test LineA set
Three-core power cordOne article
Insurance management 5ATwo branches
Qualification Certificate/Warranty Carda sheet
Packing lista sheet
Operation manual -
INSTRUCTIONS
1. The tap-changer must be reset before the reverse tap-changer is tested. After discharge, the alarm sound stops and the tap-changer can be switched.
2. Before removing wires, we must wait for the alarm to stop after the discharge, and then remove the wires.
3. When choosing the current, we should refer to the range in the technical index column, and do not use it beyond or under the range. When the range is over, the instrument is always in the "charging" state because the current can not reach the preset value. When such a state occurs, the range should be confirmed and the appropriate current should be selected for testing.
4. Pay attention to the measurement range when using the magnetic assistant method. Because two high-voltage coils are connected in parallel and one in series, the resistance of high-voltage coils is added 1.5 times in the whole test circuit, which should be included in the selection of measuring range. If the output current can't reach the set value or the output current is unstable.
5. It is a normal phenomenon that the short contacts of the three lines of the magnetizing method may have residual current when they are disconnected after discharge, and may be ignited and discharged when they are dismantled.